MAUDE Adverse Event Report: MEDACTA INTERNATIONAL SA AMISTEM C CEMENTED STD STEM SIZE1 L; CEMENTED HIP STEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDACTA INTERNATIONAL SA AMISTEM C CEMENTED STD STEM SIZE1 L; CEMENTED HIP STEM Back to Search Results
Catalog Number 01.18.151
Health Effect - Clinical Code Bone Fracture(s) (1870)
Date of Event 07/27/2016
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Batch review performed on (b)(6) 2018: lot 155668: 60 items manufactured and released on (b)(6) 2015.Expiration date: (b)(6) 2020.No anomalies found related to the issue.To date, 57 items of the same lot have been already sold without any similar reported event.
 
Event or Problem Description
On (b)(6) 2018 the sales representative has been informed by the surgeon that a revision surgery was performed.At this time we do not know if the reason of the revision surgery is bone fracture or stem loosening.
 
Additional Manufacturer Narrative
Clinical evaluation performed by medical affairs director: few months after primary hybrid tha in an elderly female patient a revision is to be performed.Reasons that led to revision surgery are not specified.Cementation technique appears to be improvable but we are not concluding that this is the cause for such an early failure.Information supplied is not sufficient to draw any conclusion.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMISTEM C CEMENTED STD STEM SIZE1 L
Common Device Name
CEMENTED HIP STEM
Manufacturer (Section D)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, 6874
SZ  6874
MDR Report Key7291629
Report Number3005180920-2018-00091
Device Sequence Number13768421
Product Code LZO
UDI-Device Identifier07630030804267
UDI-Public07630030804267
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
K103189
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2016
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/24/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number01.18.151
Device Lot Number155668
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 01/24/2018
Supplement Date Received by Manufacturer01/24/2018
Initial Report FDA Received Date02/23/2018
Supplement Report FDA Received Date04/24/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age81 YR
-
-