Catalog Number TSFB-35-260
Medical Device Problem Code Unraveled Material (1664)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 02/19/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product code = dqx.(b)(4).This report includes information known at this time. a follow-up report will be submitted should additional relevant information become available.
 
Event or Problem Description
It was reported that during a right lower extremity angiogram, the bentson straight fixed core wire guide unraveled.The wire guide was able to be pulled out of the patient, and another wire guide was utilized to successfully complete the procedure.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Additional Manufacturer Narrative
Per information received on 01mar2018, the following was updated: lot number, expiration & manufacture dates, udi.This report includes information known at this time. a follow-up report will be submitted should additional relevant information become available.
 
Additional Manufacturer Narrative
Investigation ¿ evaluation: a review of the dimensional verification, complaint history, device history record, drawing, manufacturing instructions, quality control, specifications, and a visual inspection of the returned device were conducted during the investigation.The visual inspection of the returned device confirmed the presence of several kinks near the distal end of the device.Elongation was noted from 19.8 cm to 21.1 cm from the distal end.Additionally, a document based investigation evaluation was performed.There is no evidence to suggest the product was not made to specifications.A review of the device history record shows no nonconforming events which could contribute to this failure mode.It should be noted there were no other reported complaints for this lot number.Based on the information provided, the examination of the returned product, and the results of our investigation, a definitive root cause could not be determined.We will continue to monitor for similar complaints.Per the quality engineering risk assessment, no further action is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BENTSON STRAIGHT FIXED CORE WIRE GUIDE
Common Device Name
DQX WIRE, GUIDE, CATHETER
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key7292418
Report Number1820334-2018-00558
Device Sequence Number1994713
Product Code DQX
UDI-Device Identifier00827002010635
UDI-Public(01)00827002010635(17)221115(10)8302102
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 06/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberTSFB-35-260
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 02/19/2018
Supplement Date Received by Manufacturer03/01/2018
05/30/2018
Initial Report FDA Received Date02/23/2018
Supplement Report FDA Received Date03/08/2018
06/13/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age59 YR
-
-