Catalog Number 4922250
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 01/24/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Pre-operative diagnosis for this procedure: spinal canal stenosis type of procedure or technique used : lumbar anteroposterior fusion levels implanted: l2-4 it was reported that intra-op, after inserting the cage, there was something that looked like spring, came out from the connection part between inserter and cage.It appeared to be a fragment of screw thread of the cage.No fragments remained in the patient.No patient symptoms or complications as a result of this event.
 
Additional Manufacturer Narrative
Product analysis: there is a small piece of what appears to be peek material that was returned.This could have the result of the fitment between the implant and the instrument.Since neither was returned no further analysis can be done.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLYDESDALE PTC SPINAL SYSTEM
Common Device Name
INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, LUMBAR
Manufacturer (Section D)
MSD DEGGENDORF MFG
wertstrasse 17
deggendorf 94469
MDR Report Key7292425
Report Number1030489-2018-00278
Device Sequence Number1177619
Product Code MAX
UDI-Device Identifier00643169191327
UDI-Public00643169191327
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K133205
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/04/2024
Device Catalogue Number4922250
Device Lot Number66CT
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/16/2018
Initial Date Received by Manufacturer 01/24/2018
Supplement Date Received by Manufacturer04/23/2018
Initial Report FDA Received Date02/23/2018
Supplement Report FDA Received Date05/22/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age78 YR
Patient Weight60
-
-