• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SYNTHES PRODUCTS LLC SMALL BATTERY DRIVE; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES PRODUCTS LLC SMALL BATTERY DRIVE; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT Back to Search Results
Catalog Number 532.001
Device Problems Device Stops Intermittently (1599); Mechanical Jam (2983)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/22/2018
Event Type  malfunction  
Manufacturer Narrative
The device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.During repair, it was determined that the trigger was blocked, jammed and moving heavily.Therefore, the reported condition was confirmed.However, the assignable root cause could not be determined.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Event Description
This is report 1 of 2 for the same event.It was reported from (b)(6) that that two small battery drive device would sometimes work and sometimes would not work.It was not reported if the device was used in surgery, or if there was patient involvement.It was not reported if there were any delays in a surgical procedure or if a spare device was available.It was not reported if there were any injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Manufacturer Narrative
Device evaluation update: further investigation determined that the assignable root cause was determined to be due to premature wear.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SMALL BATTERY DRIVE
Type of Device
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
SYNTHES PRODUKTIONS GMBH
hauptstrasse 24
waldenburg 4437
SZ   4437
Manufacturer Contact
michael cote
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key7292476
MDR Text Key101006792
Report Number8030965-2018-51320
Device Sequence Number1
Product Code HWE
UDI-Device Identifier7611819107083
UDI-Public(01)7611819107083(11)120807
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/28/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/23/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number532.001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/05/2018
Is the Reporter a Health Professional? No
Date Manufacturer Received04/02/2018
Was Device Evaluated by Manufacturer? No
Date Device Manufactured08/07/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-