U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SYNTHES PRODUCTS LLC SMALL BATTERY DRIVE INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES PRODUCTS LLC SMALL BATTERY DRIVE INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT Back to Search Results
Catalog Number 532.001
Device Problems Device Stops Intermittently (1599); Mechanical Jam (2983)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/22/2018
Event Type  malfunction  
Manufacturer Narrative
Device was used for treatment, not diagnosis. The device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment. During repair, it was determined that the trigger was blocked, jammed and moving heavily. Therefore, the reported condition was confirmed. The assignable root cause was determined to be due to premature wear. If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Event Description
This is report 2 of 2 for the same event it was reported from (b)(6) that that two small battery drive device would sometimes work and sometimes would not work. It was not reported if the device was used in surgery, or if there was patient involvement. It was not reported if there were any delays in a surgical procedure or if a spare device was available. It was not reported if there were any injuries, medical intervention or prolonged hospitalization. All available information has been disclosed. If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameSMALL BATTERY DRIVE
Type of DeviceINSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
eimattstrasse 3
oberdorf 4436
SZ 4436
Manufacturer (Section G)
SYNTHES PRODUKTIONS GMBH
hauptstrasse 24
waldenburg 4437
SZ 4437
Manufacturer Contact
michael cote
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key7292481
MDR Text Key100913458
Report Number8030965-2018-51321
Device Sequence Number1
Product Code HWE
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation
Type of Report Initial
Report Date 01/28/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received02/23/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Catalogue Number532.001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/05/2018
Is the Reporter a Health Professional?
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received01/28/2018
Was Device Evaluated by Manufacturer? No Answer Provided
Date Device Manufactured08/07/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unkown

-
-