X-ray analysis of the valve confirmed the presence of extensive intrinsic calcification in all the leaflets.Histological examination revealed that the stent and sewing ring were covered by fibrous pannus on the inflow side, which protruded 2 mm into the lumen, narrowing the annulus and contributing to stenosis.The pericardium of all leaflets were thickened and stiffened due to intrinsic and/or vegetating calcification, which further contributed to the prosthetic stenosis and subsequent insufficiency due to tears on all leaflets.Blood entrapment was visible on all surfaces.No bacteria were detected with gram staining.The manufacturing and material records for the mitroflow bioprosthetic pericardial heart valve, model # dla23, s/n # (b)(4), as they pertain to the reported event, were retrieved and reviewed by quality engineering at livanova (b)(4).The results confirmed that this valve satisfied all material, visual, and performance standards required for a dla23 mitroflow aortic pericardial heart valve at the time of manufacture and release.Based on the analyses performed, leaflet calcification caused stiffening and led to progressive valve stenosis.Pannus overgrowth also contributed to the observed stenosis.Structural dysfunction is a major cause of failure of bioprosthetic heart valves, and the principle underlying cause is cuspal calcification.As limited patient history was provided, it is possible that the patient's clinical conditions may have contributed to the structural valve deterioration observed in the valve; however, the root cause of the event is ultimately unknown.
|
|
A mitroflow dla was explanted on (b)(6) 2018, 4.47 years after implant due to svd, and a trifecta size 21 valve was implanted in its place.The patient recovered without incident.No risk factors for svd were reported.There were no episodes of clinical or subclinical endocarditis, and specimens from surgery were all negative (including pcr amplification).
|