It was reported that during a procedure for implanting a synthes lateral entry recon nail on (b)(6) 2018, a 4.5mm/6.5mm stepped drill bit large qc/485mm would not pass through the protection sleeves and the nail as it should.During this process, the drill bit broke inside of the patient bone.The broken fragment of the drill bit was retrieved and nothing was left in the patient.Also, a recon locking aiming arm for lateral entry femoral nails-ex was not working properly as it was misaligned with nail, but prior to insertion of femoral neck/head component, nail connection was checked and the aiming arm connection was secured as well.Two of the 11.5mm/8.5mm protection sleeve for recon locking and two 8.5mm/3.2mm wire guide for recon locking were also not guiding the drill bit properly through the nail.A standard insertion handle connected with the aiming arm wasn't working properly as the sales consultant heard some noise when the handle was being used.A cannulated connecting screw for standard insertion handle was not getting connected properly and a 5.0mm flexible shaft was not able to pass over a 2.5mm reaming rod.There was a surgical delay of 30 minutes and the surgeon was unable to implant recon screws into proximal femur.To complete the surgery, the surgeon opted to lock in dynamic hole by using a regular 4.9mm locking volt.The surgery was completed and the patient was stable after the surgery.This complaint involves six (6) devices.This report is 2 of 6 for (b)(4).
|
|
Customer quality conducted an investigation of the returned device.Service & repair evaluation the customer reported the instrument was not aligning correctly with the nail.The repair technician reported the instrument had no issues, but could not be completely evaluated for misalignment.Due to wear the device was deemed not repairable and forwarded to customer quality.Device condition: the aiming arm was received at cq intact with surface wear consistent with use and which would not impact the functionality.The device was functionally tested with the returned mating protection sleeves, 03.010.075, and found to function as intended.The sleeves could be fully advanced and retracted and the locking cams retained and released the sleeves as intended.The aiming arm was also functionally tested with the returned mating insertion handle, 03.010.045.The device fully assembled and disassembled as intended.Functional testing with the nail could not be performed as the mating nail was not received.The device was inspected in the ¿drawing review¿ section below and no discrepancies were observed.Thus, as no functional or dimensional issue was identified, the complaint condition for this device is unconfirmed and could not be replicated.Lot number review: review of the device history record(s) showed that there were no issues during the manufacture of the product that would contribute to this complaint condition.The lot went through the required steps during the inspection at the time of manufacturing and the dhr records showed no issues concerning the material or material conditioning.Release to warehouse date: 29 nov 2007 drawing review: based on the date of manufacture the drawings reflecting the current and manufactured revision, were reviewed.During the investigation no product design issues or discrepancies were observed that may have contributed to the complaint condition.The recon locking holes were inspected and found to both measure 11.5mm.This is within the specification of 11.5mm +.036/-0 per drawing.The static and dynamic locking holes were inspected and found to both measure 12.0mm.This is within the specification of 12mm +.036/-0 per drawing.No definitive root cause was able to be determined as no product issues were identified with the aiming arm.The insertion handle 03.010.076, which is also a part of the alignment construct, had a broken tab.This tab is responsible for alignment and thus, would impact the ability to align the construct.Conclusion: a device history record (dhr) review, device inspection, and drawing review were performed as part of this investigation.The complaint condition is unconfirmed as no issues were identified with the returned device.There is no indication that a design or manufacturing issue contributed to the complaint.No new malfunctions were observed during the course of this investigation.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Additional narrative: corrected data: these dates were reported as (b)(6) 2018 in the initial report (b)(4) in error.The correct date is (b)(6) 2018.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|