• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. POD5; HCG, KRD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. POD5; HCG, KRD Back to Search Results
Catalog Number RBYPOD5
Medical Device Problem Codes Failure to Advance (2524); Physical Resistance (2578)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/25/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product was disposed of by the hospital and is no longer available for return.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure in the right internal iliac artery (iia) using pod5's.It was noted that the patient's anatomy was slightly tortuous and calcified.During the procedure, the physician deployed and detached a pod coil in the target vessel using a lantern delivery microcatheter (lantern).While attempting to advance a pod5 through the lantern, the physician experienced resistance around the hub of the lantern.The physician then experienced resistance again after advancing the pod5 approximately 10 centimeters into the lantern; therefore, the pod5 was removed.The pod5 was then soaked in saline and the lantern flushed through with saline.While attempting to re-advance the pod5 through the lantern, resistance was encountered again at the same position; therefore, it was removed.The procedure was completed using the same lantern, a new pod5 and additional coils.There was no report of an adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POD5
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key7293763
Report Number3005168196-2018-00407
Device Sequence Number956654
Product Code HCG
UDI-Device Identifier00814548013787
UDI-Public00814548013787
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K141134
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign,health pr
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 01/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date09/04/2024
Device Catalogue NumberRBYPOD5
Device Lot NumberF71430
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/26/2018
Initial Report FDA Received Date02/23/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age74 YR
Patient Weight70
-
-