| Model Number |
MICL13.2 |
| Medical Device Problem Code |
Device Operates Differently Than Expected (2913)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
01/23/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
No similar complaints types events were reported for units within the same lot.Lens was returned in liquid, in lens vial.Visual inspection found no visible damage to the lens.(b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter indicated that during loading of a micl13.2, -15.0 diopter, implantable collamer lens, it was noted that a portion of one of the footplates was missing.There was no patient contact.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected data: expiration date: it is corrected from "31-jan-2018" to "31-jan-2019".Claim# (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected data: device evaluation: lens was returned in liquid inside a lens vial.Visual inspection found the lens missing a piece of its haptic.Claim# (b)(4).
|
| |
|