• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL OPTICS TECNIS; MULTIFOCAL IOLS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL OPTICS TECNIS; MULTIFOCAL IOLS Back to Search Results
Model Number ZMA00
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Blurred Vision (2137); No Code Available (3191)
Date of Event 02/12/2013
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The intraocular lens remains implanted.(b)(4).Device evaluation: the lens remains implanted; therefore, it was not returned for investigation.The reported complaint was not verified.Manufacturing records review: the manufacturing records for the product were reviewed.The product was manufactured and released according to specifications.A search revealed that no additional complaints for this order number have been received.Labeling review: the directions for use (dfu) were reviewed.The dfu provide the customer with proper usage instructions and guidelines.Conclusion: as a result of the investigation, there is no indication of a product quality deficiency.All pertinent information available to abbott medical optics has been submitted.
 
Event or Problem Description
It was reported that a patient had multifocal lenses implanted bilaterally on (b)(6) 2013.She has had issues ever since.She cannot see especially at night.She had laser done to correct for astigmatism on (b)(6) 2013.She reported that it is better but still not the best and described her vision as cloudy.She had bilateral yag capsulotomy done on (b)(6) 2013 and again in the right eye in 2016 to treat significant posterior capsule opacification (pco).The patient mentioned that she is worried about posterior chamber weakening.The last surgeon she went to see (dr.(b)(6)) suggested an explant but the patient was concerned about explanting the lens after this many years.During follow up with the doctor he mentioned that he did not notice any opacification on the intraocular lenses, but there were posterior capsular opacities.It was mentioned that the patient had increased blurring in left eye for the past 2 months.The doctor noticed that when he checked her eyes this month again, he noticed significant pco on the residual posterior capsule affecting a large portion of the rings in the optic of the lens in the left eye.The patient is seeking further opinion on iol exchange.The doctor plans to perform supplementary yag post capsulotomy on the left eye if the patient does not plan to have iol exchange.No additional information was provided to abbott medical optics.Note: 2 mdr¿s will be submitted, one for each eye.This report pertains to the right eye.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS
Common Device Name
MULTIFOCAL IOLS
Manufacturer (Section D)
ABBOTT MEDICAL OPTICS
santa ana CA
MDR Report Key7294492
Report Number2648035-2018-00291
Device Sequence Number1211705
Product Code MFK
UDI-Device Identifier05050474536692
UDI-Public(01)05050474536692(17)140804
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
P080010
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Reporter Type Manufacturer
Report Source company representative,consum
Initial Reporter Occupation Patient
Type of Report Initial
Report Date (Section B) 02/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Physician
Device Expiration Date08/04/2014
Device Model NumberZMA00
Device Catalogue NumberZMA00U0145
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/26/2018
Initial Report FDA Received Date02/24/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age53 YR
-
-