Model Number CC4204A
Medical Device Problem Codes Device Operates Differently Than Expected (2913); Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Capsular Bag Tear (2639); Vitrectomy (2643)
Date of Event 12/18/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Collamer ultraviolet-absorbing posterior chamber single piece foldable intraocular lens.Work order search: no similar complaint was reported for units within the same lot.(b)(4).
 
Event or Problem Description
The reporter indicated the surgeon inserted a cc4204a collamer single piece lens, +17.5 diopter, in the patient's right (od) eye.Upon insertion, the posterior capsule was torn.The lens was cut in half to remove from the eye within the same surgery, with no patient injury.A vitrectomy was performed and an anterior chamber lens was implanted.The reporter stated the cause of the event was unknown.The lens was discarded and will not be returned.
 
Additional Manufacturer Narrative
(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SEE H-10
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
MDR Report Key7294654
Report Number2023826-2018-00277
Device Sequence Number957332
Product Code HQL
UDI-Device Identifier00841542102346
UDI-Public00841542102346
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
P990013
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2017
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 01/31/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Physician
Device Expiration Date01/31/2018
Device Model NumberCC4204A
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 01/31/2018
Supplement Date Received by Manufacturer04/27/2018
Initial Report FDA Received Date02/24/2018
Supplement Report FDA Received Date05/03/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age73 YR
-
-