Model Number TA3048L
Medical Device Problem Code Failure to Fire (2610)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 02/06/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, intra-operatively, the trigger did not load and did not fire at the time of stapling.There was no patient injury as reported.
 
Additional Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of device.Visual inspection of the single use loading unit (sulu) cartridge noted that a full staple complement was present.The clinical instrument was not received.Since the clinical instrument was not received a pmv representative instrument was used for testing.Functionally, the representative instrument and sulu were applied to test media with proper staple formation.Although the reported condition was not confirmed, it may occur when the user mistakes the tactile pre-clamp feature for the full firing stroke.This feature allows the handle to be released, without returning to its original position, for final positioning of the tissue.After the tissue is properly positioned, the handle stroke must be continued to full clamp position.Staples will only fire after the fully clamped handle returns to its original position and is squeezed a second time.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.The investigation concluded there were no abnormalities that would have caused or contributed to the reported condition.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, intra-operatively, the trigger did not load and did not fire at the time of stapling.A full squeeze of the device was performed.There was no patient injury as reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TA
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
MDR Report Key7296555
Report Number2647580-2018-01009
Device Sequence Number11133769
Product Code GDW
UDI-Device Identifier10884521043121
UDI-Public10884521043121
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K111825
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date (Section B) 11/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2018
Device Model NumberTA3048L
Device Catalogue NumberTA3048L
Device Lot NumberP3J0175X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/26/2018
Initial Date Received by Manufacturer 02/08/2018
Supplement Date Received by Manufacturer10/29/2018
Initial Report FDA Received Date02/26/2018
Supplement Report FDA Received Date11/20/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-