• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE DIAGNOSTICS INC. DIMENSION VISTA®; DIMENSION VISTA® PHOSPHORUS FLEX® REAGENT CARTRIDGE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE DIAGNOSTICS INC. DIMENSION VISTA®; DIMENSION VISTA® PHOSPHORUS FLEX® REAGENT CARTRIDGE Back to Search Results
Catalog Number K1061 SMN10445144
Medical Device Problem Code Low Test Results (2458)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 01/29/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The customer contacted the siemens customer care center (ccc) for the discordant phosphorus (phos) results obtained on the dimension vista instrument.Siemens healthcare diagnostics headquarters support center (hsc) concluded their investigation of the discordant phosphorus (phos) results.Headquarters support center (hsc) has reviewed the information provided as well as a review of instrument data.Based on review of calibration history, the customer calibrated phos using reagent lot 17291ac and used this calibration until (b)(6) 2018 when qc recovery was out of range high.The qc was run using flex sequence 104, well set 6;10;2.The issue was resolved by recalibrating phos using a new flex sequence 103 and well set 5;9;1.The discrepant patients were repeated and corrected reports were issued.After recalibration, the instrument is currently working as specified.No further action will be taken by hsc.No further evaluation of the device is required.
 
Event or Problem Description
Discordant depressed phosphorus (phos) results were obtained on multiple patient samples on the dimension vista 1500 instrument.The initial results were reported.One patient received treatment based on the discrepant result.The same patient samples were repeated after recalibration of the same flex reagent cartridge with an alternate flex well set higher results were obtained.Corrected reports were issued.There are no reports of patient intervention or adverse health consequences due to the discordant depressed phos results or due to patient treatment.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DIMENSION VISTA®
Common Device Name
DIMENSION VISTA® PHOSPHORUS FLEX® REAGENT CARTRIDGE
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS INC.
500 gbc drive
po box 6101
newark DE 19714 6101
MDR Report Key7297195
Report Number2517506-2018-00102
Device Sequence Number11098329
Product Code CEO
UDI-Device Identifier00842768015700
UDI-Public0842768015700
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K061655
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Health Professional
Type of Report Initial
Report Date (Section B) 02/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/18/2018
Device Catalogue NumberK1061 SMN10445144
Device Lot Number17291AC
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/31/2018
Initial Report FDA Received Date02/26/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-