Dexcom was made aware on (b)(4) 2018 that on (b)(6) 2018, there was an inaccuracy between the continuous glucose monitor (cgm) and the blood glucose (bg) meter.The sensor was inserted into the abdomen on (b)(6) 2018.Reportedly, the patient used an expired sensor.No additional event or patient information is available.No data was provided for evaluation.The complaint confirmation and probable cause could not be determined.Labeling indicates: do not insert sensors past the expiration date.The expiration date format is yyyy-mm-dd.Insert sensors on or before the end of the calendar day printed on the sensor package label.The complaint device was returned for evaluation.The device was visually inspected and no defect was found.Functional testing was performed and the reported fault could not be reproduced and there was no failure detected.A voltage test was performed and it passed.The reported event of inaccuracies was not confirmed.The probable root cause cannot be determined.
|