| Medical Device Problem Code |
Structural Problem (2506)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The pump has not been returned to animas for evaluation.If the device is returned, an evaluation shall be completed and a supplemental report will be filed.No conclusions can be made at this time.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2018, the reporter contacted animas, alleging a casing/condition (damaged cap and case) issue.The reporter alleged the battery compartment cap will not attach, and there is a crack in the battery compartment.There was no indication that the product caused or contributed to an adverse event.This complaint is being reported because the alleged malfunction has the ability to result in a delay in treatment or long term cessation in delivery if the damage impacts the power circuit.
|
| |
|
Additional Manufacturer Narrative
|
|
Follow-up #1: date of submission (b)(6) 2018 device evaluation: the device has been returned and evaluated by product analysis on (b)(6) 2018 with the following findings: during investigation, the battery compartment was cracked from the threads to the case seal left of the grip pad.The battery cap threads were damaged and difficult to remove from the test pump.Unrelated to the original complaint, moisture was observed in the battery compartment with a leak.The pump was opened to find no moisture.
|
| |
|