| Model Number |
METRIX AIR |
| Medical Device Problem Code |
Incorrect Or Inadequate Test Results (2456)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
01/31/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Internal report # (b)(4).Returned meter evaluated with no defect found.Test strips not returned for evaluation.Most likely underlying root cause: mlc-20-user's test strip had poor storage (kitchen).Test strip udi# (b)(4).Note: manufacturer contacted customer on (b)(6) 2018 in a follow-up call in order to ensure that the replacement products resolved the initial concern; customer stated that the replacement product is working to their satisfaction and they have not had any medical intervention since the last call.
|
| |
|
Event or Problem Description
|
|
Consumer reported complaint for erratic blood glucose results.Spouse is calling on behalf of the customer.The customer is concerned with results obtained of 106 and 266mg/dl.The expected fasting blood glucose test result range is 115 to 140mg/dl.The customer feels well and did not report symptoms.Medical attention is not reported as a result of the actual blood glucose results.The product is not stored by customer according to specification in the kitchen.The test strip lot manufacturer's expiration date is 1/21/2019 and open vial date is less than 4 months ago.The meter memory was reviewed for previous test result history: (date/time not stored correctly) (b)(6).Test were made on the same hand, different finger.Unknown if results are fasting or non-fasting.
|
| |
|
Search Alerts/Recalls
|