• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TRIVIDIA HEALTH, INC. METRIX AIR; BLOOD GLUCOSE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TRIVIDIA HEALTH, INC. METRIX AIR; BLOOD GLUCOSE SYSTEM Back to Search Results
Model Number METRIX AIR
Medical Device Problem Code Incorrect Or Inadequate Test Results (2456)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 01/31/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Internal report # (b)(4).Returned meter evaluated with no defect found.Test strips not returned for evaluation.Most likely underlying root cause: mlc-20-user's test strip had poor storage (kitchen).Test strip udi# (b)(4).Note: manufacturer contacted customer on (b)(6) 2018 in a follow-up call in order to ensure that the replacement products resolved the initial concern; customer stated that the replacement product is working to their satisfaction and they have not had any medical intervention since the last call.
 
Event or Problem Description
Consumer reported complaint for erratic blood glucose results.Spouse is calling on behalf of the customer.The customer is concerned with results obtained of 106 and 266mg/dl.The expected fasting blood glucose test result range is 115 to 140mg/dl.The customer feels well and did not report symptoms.Medical attention is not reported as a result of the actual blood glucose results.The product is not stored by customer according to specification in the kitchen.The test strip lot manufacturer's expiration date is 1/21/2019 and open vial date is less than 4 months ago.The meter memory was reviewed for previous test result history: (date/time not stored correctly) (b)(6).Test were made on the same hand, different finger.Unknown if results are fasting or non-fasting.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
METRIX AIR
Common Device Name
BLOOD GLUCOSE SYSTEM
Manufacturer (Section D)
TRIVIDIA HEALTH, INC.
2400 nw 55th court
fort lauderdale FL 33309
MDR Report Key7297872
Report Number1000113657-2018-00233
Device Sequence Number13762229
Product Code NBW
UDI-Device Identifier00021292007911
UDI-Public(01)00021292007911
Combination Product (Y/N)N
Initial Reporter StateWA
PMA/510(K) Number
K150052
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Remedial Action Replace
Type of Report Initial
Report Date (Section B) 02/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMETRIX AIR
Device Catalogue NumberREA4H01-01
Device Lot NumberMU2559
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/22/2018
Was the Report Sent to FDA? No
Event Location Home
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/31/2018
Initial Report FDA Received Date02/26/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-