| Model Number |
48080230 |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
02/01/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the snake arm would not latch onto decompression tubes.There was a reported 15 minute delay.The procedure was successfully completed, and there were no reports of adverse consequences.
|
| |
|
Additional Manufacturer Narrative
|
|
Method: visual inspection, functional inspection, device history review, complaint history review, risk assessment; result: the customer reported event was not confirmed as the device was functional during inspection.Product inspection upon return showed no deformation on the instruments.Manufacturing records were reviewed for the corresponding lot and no relevant issues were identified.Conclusion: since the device was found to the fully functional, the reported issue could not be confirmed.
|
| |
|
Event or Problem Description
|
|
It was reported that the snake arm would not latch onto decompression tubes.There was a reported 15 minute delay.The procedure was successfully completed, and there were no reports of adverse consequences.
|
| |
|