• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. VACCESS PTA BALLOON DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. VACCESS PTA BALLOON DILATATION CATHETER Back to Search Results
Catalog Number VA8078
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 12/29/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No medical records and no medical images were provided to the manufacturer.The lot number for the device was provided.The device history records are currently under review.The return of the device is pending.The investigation is currently under way.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that the pta balloon catheter allegedly ruptured at 4 atm.There was no reported retraction difficulty through the sheath.Reportedly, another balloon catheter was exchanged over the guidewire and the procedure was completed.There was no reported patient injury.
 
Additional Manufacturer Narrative
Manufacturing review: the lot number was provided and the lot device history records have been reviewed.The lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the investigation is inconclusive for the reported rupture, as the device was not returned for evaluation.The definitive root cause for the reported rupture could not be determined based upon the available information.It is unknown if patient and/or procedural issues contributed to the reported event.Labeling review: the current ifu (instructions for use) states: warnings: when the catheter is exposed to the vascular system, it should be manipulated while under high-quality fluoroscopic observation.Do not advance or retract the catheter unless the balloon is fully deflated.If resistance is met during manipulation, determine the cause of the resistance before proceeding.Applying excessive force to the catheter can result in tip breakage or balloon separation.Do not exceed the rbp recommended for this device.Balloon rupture may occur if the rbp rating is exceeded.To prevent over pressurization, use of a pressure monitoring device is recommended.Potential adverse reactions: additional intervention.
 
Event or Problem Description
It was reported that the pta balloon catheter allegedly ruptured at 4 atm.There was no reported retraction difficulty through the sheath.Reportedly, another balloon catheter was exchanged over the guidewire and the procedure was completed.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VACCESS PTA BALLOON DILATATION CATHETER
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key7298185
Report Number2020394-2018-00155
Device Sequence Number16514875
Product Code DQY
UDI-Device Identifier00801741059933
UDI-Public(01)00801741059933
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K111850
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/12/2019
Device Catalogue NumberVA8078
Device Lot Number93ZA0073
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 01/31/2018
Supplement Date Received by Manufacturer03/16/2018
Initial Report FDA Received Date02/26/2018
Supplement Report FDA Received Date04/10/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-