• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALERE SAN DIEGO, INC. RED -MLTI-DRG 6 PNL SCRN TST (DOA-164); ENZYME IMMUNOASSAY, COCAINE AND COCAINE METABOLITES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALERE SAN DIEGO, INC. RED -MLTI-DRG 6 PNL SCRN TST (DOA-164); ENZYME IMMUNOASSAY, COCAINE AND COCAINE METABOLITES Back to Search Results
Model Number DOA-164
Medical Device Problem Code False Positive Result (1227)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation conclusion: the customer's observation was not replicated in-house with retention products.Retention devices were tested with in-house drug-free donor urine and all devices showed expected negative results; all coc, met, thc, mop, oxy, and bzo strips yielded correct negative results at read time.No false positive results were observed during in-house testing.Manufacturing batch record review did not uncover any abnormalities.The device provides only a qualitative, preliminary analytical result.A secondary analytical method must be used to obtain a confirmed result.Based on the information available, there is no indication of a product deficiency.A capa, capa (b)(4), has been opened to address the late escalation of this complaint.
 
Event or Problem Description
The customer reported receiving unconfirmed false positive results for all analytes using the multi-drug one step 6 drug screen test panel.The unconfirmed false positive results occurred on different random donors.The unconfirmed false positive results occurred on 23 devices out of 25 devices used.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RED -MLTI-DRG 6 PNL SCRN TST (DOA-164)
Common Device Name
ENZYME IMMUNOASSAY, COCAINE AND COCAINE METABOLITES
Manufacturer (Section D)
ALERE SAN DIEGO, INC.
9975 summers ridge road
san diego CA 92121
MDR Report Key7298223
Report Number2027969-2018-00025
Device Sequence Number13765683
Product Code DIO
UDI-Device Identifier10858904006317
UDI-Public(01)10858904006317(17)180831(10)DOA6090545
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K061718
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer,health professional
Initial Reporter Occupation Health Professional
Type of Report Initial
Report Date (Section B) 02/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2018
Device Model NumberDOA-164
Device Lot NumberDOA6090545
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/09/2018
Initial Report FDA Received Date02/26/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-