| Model Number |
3418 |
| Medical Device Problem Code |
Incorrect Measurement (1383)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/31/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Multiple attempts for product return and requests for additional information were made.The product has not been returned to masimo to allow an analysis to be performed.If new information is obtained or the product is returned, a follow up report will be submitted.,.
|
| |
|
Event or Problem Description
|
|
The customer reported they are seeing a three point difference in the sphb reading from the pronto and a blood draw.No patient impact or consequences were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The returned sensor was evaluated.During evaluation the sensor passed all visual and functional testing.The sensor was determined to be functioning as designed.Corrected data: lot #: updated from "a17e130" to "17gd2." device manufacture date updated from "07/06/2017" to "07/25/2017.".
|
| |
|