| Model Number |
8225101 |
| Medical Device Problem Code |
Device Operates Differently Than Expected (2913)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/27/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Evaluation was not performed; product was not returned for analysis.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
The customer reported that the standard prass probe stimular did not respond at all.No patient impact or injury.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Additional information indicated that the that the stim return cable was plugged into the wrong port (stim 2) and this is why this problem occurred.The nim was set at 1.0, device had wave forms from usual manipulation (little blips), green ticks were obtained; white line present.The probe was discarded.
|
| |
|