• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS, INC. 1818910 SIGMA POR TRAY SCREW 6.5/35MM; BONE SCREWS AND PINS : KNEE TIBIAL TRAY ACCESSORY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS, INC. 1818910 SIGMA POR TRAY SCREW 6.5/35MM; BONE SCREWS AND PINS : KNEE TIBIAL TRAY ACCESSORY Back to Search Results
Catalog Number 875704
Medical Device Problem Code Failure To Adhere Or Bond (1031)
Health Effect - Clinical Code Pain (1994)
Date of Event 02/07/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
The patient was revised to address knee pain in the total knee.It was stated that, there was no patella in the patient from the first surgery.Bone scan was hot around the tibia.It was decided by the doctor that the tibia was loose at the bone to implant interface.Pressfit sigma fb tibia was pulled out and a sigma fixed bearing cemented tibia with a stem replaced the pressfit one.The patella was resurfaced at this time as well.Doi: (b)(6) 2016.Dor: (b)(6) 2018.Right knee.
 
Additional Manufacturer Narrative
(b)(4).Investigation summary: no device associated with this report was received for examination.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SIGMA POR TRAY SCREW 6.5/35MM
Common Device Name
BONE SCREWS AND PINS : KNEE TIBIAL TRAY ACCESSORY
Manufacturer (Section D)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic drive
warsaw IN 46582 0988
MDR Report Key7298832
Report Number1818910-2018-54133
Device Sequence Number877387
Product Code LPH
UDI-Device Identifier10603295218685
UDI-Public10603295218685
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K983014
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date (Section B) 02/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/31/2018
Device Catalogue Number875704
Device Lot Number475009
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 02/07/2018
Supplement Date Received by Manufacturer11/13/2018
Initial Report FDA Received Date02/27/2018
Supplement Report FDA Received Date11/14/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age74 YR
Patient Weight97
-
-