510(k) number: k160229.Cook ireland ltd (manufacturer) is submitting this report on behalf of cook medical incorporated (cmi)(importer).Exemption number: e2016031.Information pertaining to section g.1 as follows: importer site contact and address: (b)(4).Cook medical incorporated (cmi), 1025 acuff road, p.O box 4195, bloomington , indiana 47402-4195.Importer site establishment registration number: 3005580113.Problem statement: "after vocal cords and subsegmental bronchus was observed by endoscopy, ebus-tbna (endoscopic ultrasound guided transbronchial needle aspiration) was performed.The user inserted and positioned olympus' bf-uc260fw at 4r (right lower paratracheal nodes).Then he inserted echo-hd-25-ebus-o-c with the stylet inserted fully and punctured the target site.It was confirmed that the some cells was collected successfully at first puncture.Then, the user attempted second puncture, however the needle did not stick in because the puncture quality of the needle was not good, so he removed the needle from the endoscope and it was confirmed that the needle bent at 2~3cm from the tip.Some cells were collected by the first puncture, so he completed the procedure without using other devices to collect more cells.[note] the first puncture was smooth.At the second puncture, the needle was inserted into the scope smoothly, but it id not stick well.I cannot deny the possibility that the needle penetrated cartilages, or advanced the surface of cartilage." lab evaluation: 1 x echo-hd-25ebus-0-c device of lot # c1384037 was returned to cirl for evaluation.Upon evaluation, the device was returned in a plastic bag.There was no stylet returned, no needle exposure.The needle tip was bent distally.The needle advances and retracts without any issue.The needle tip is intact.The device was flushed with saline.There is no biological matter present on the needle.There are no other functional issue with the device.There was a bend below the sheath extender but this can be attributed to the conditions in which the device was returned.The customer complaint is confirmed as there was a distal kink noted on the needle during the lab.There was a bend below the sheath extender this can be attributed to the condition in which the device was returned.Root cause: a definitive root cause for the customer complaint could not be determined as the exact operational conditions of use could not be replicated in the laboratory setting.However it is possible that the needle bent when excessive force was applied possibly while advancing the needle into the target site, also the user also stated " i cannot deny the possibility that the needle penetrated cartilages, or advanced the surface of cartilage.".Documents review: prior to distribution, all echo-hd-25-ebus-0-c devices are subjected to functional checks and visual inspection to ensure device integrity.These inspections and functional checks are outlined in internal procedures in place at cirl.As per the finished product qc instruction, the mtm is instructed to ¿measure the needle extension with the outer handle positioned at reference mark 5 and with the sheath in a straight position.The needle extension should be as per the drawing and free from any damage / kinks.¿ all products and packaging are 100% inspected for visual irregularities such as holes, dents, kinks or tears.Ifu review: it may be noted that according to the instructions for use, the user is instructed to: ¿visually inspect with particular attention to kinks, bends and breaks.If any abnormality is detected that would prohibit proper working condition, do not use¿.On review of the information provided, there is no viable evidence to suggest that the user did not follow the instructions for use.A review of the manufacturing records for echo-hd-25-ebus-0-c of lot number c1384037 did not reveal any discrepancies that could have contributed to this complaint issue.There is no evidence to suggest that this issue affects the entire lot # c1384037 upon review of complaints this failure mode has occurred previously with this lot # c1384037.Please see e-mail from senior quality engineer for justification.Summary: from the information provided, no adverse effects to the patient have been reported as a result of this occurrence.Complaints of this nature will continue to be monitored for potential emerging trends.H3 other text : 510(k) number: k160229.Cook ireland ltd (manufacturer) is submitting this report on behalf of cook medical incorporated (cmi)(importer).Exemption number: e2016031.Information pertaining to section g.1 as follows: importer site contact and address: (b)(4).Cook medical incorporated (cmi), 1025 acuff road, p.O box 4195, bloomington , indiana 47402-4195.Importer site establishment registration number: 3005580113.Problem statement: "after vocal cords and subsegmental bronchus was observed by endoscopy, ebus-tbna (endoscopic ultrasound guided transbronchial needle aspiration) was performed.The user inserted and positioned olympus' bf-uc260fw at 4r (right lower paratracheal nodes).Then he inserted echo-hd-25-ebus-o-c with the stylet inserted fully and punctured the target site.It was confirmed that the some cells was collected successfully at first puncture.Then, the user attempted second puncture, however the needle did not stick in because the puncture quality of the needle was not good, so he removed the needle from the endoscope and it was confirmed that the needle bent at 2~3cm from the tip.Some cells were collected by the first puncture, so he completed the procedure without using other devices to collect more cells.[note] the first puncture was smooth.At the second puncture, the needle was inserted into the scope smoothly, but it id not stick well.I cannot deny the possibility that the needle penetrated cartilages, or advanced the surface of cartilage." lab evaluation: 1 x echo-hd-25ebus-0-c device of lot # c1384037 was returned to cirl for evaluation.Upon evaluation, the device was returned in a plastic bag.There was no stylet returned, no needle exposure.The needle tip was bent distally.The needle advances and retracts without any issue.The needle tip is intact.The device was flushed with saline.There is no biological matter present on the needle.There are no other functional issue with the device.There was a bend below the sheath extender but this can be attributed to the conditions in which the device was returned.The customer complaint is confirmed as there was a distal kink noted on the needle during the lab.There was a bend below the sheath extender this can be attributed to the condition in which the device was returned.Root cause: a definitive root cause for the customer complaint could not be determined as the exact operational conditions of use could not be replicated in the laboratory setting.However it is possible that the needle bent when excessive force was applied possibly while advancing the needle into the target site, also the user also stated " i cannot deny the possibility that the needle penetrated cartilages, or advanced the surface of cartilage.".Documents review: prior to distribution, all echo-hd-25-ebus-0-c devices are subjected to functional checks and visual inspection to ensure device integrity.These inspections and functional checks are outlined in internal procedures in place at cirl.As per the finished product qc instruction, the mtm is instructed to ¿measure the needle extension with the outer handle positioned at reference mark 5 and with the sheath in a straight position.The needle extension should be as per the drawing and free from any damage / kinks.¿ all products and packaging are 100% inspected for visual irregularities such as holes, dents, kinks or tears.Ifu review: it may be noted that according to the instructions for use, the user is instructed to: ¿visually inspect with particular attention to kinks, bends and breaks.If any abnormality is detected that would prohibit proper working condition, do not use¿.On review of the information provided, there is no viable evidence to suggest that the user did not follow the instructions for use.A review of the manufacturing records for echo-hd-25-ebus-0-c of lot number c1384037 did not reveal any discrepancies that could have contributed to this complaint issue.There is no evidence to suggest that this issue affects the entire lot # c1384037 upon review of complaints this failure mode has occurred previously with this lot # c1384037.Please see e-mail from senior quality engineer for justification.Summary: from the information provided, no adverse effects to the patient have been reported as a result of this occurrence.Complaints of this nature will continue to be monitored for potential emerging trends.
|