• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CODMAN & SHURTLEFF, INC. / MEDOS S.A. HAKIM PROGRAMMABLE VALVE WITH PRECHAMBER; SHUNT, CENTRAL NERVOUS SYSTEM & COMPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CODMAN & SHURTLEFF, INC. / MEDOS S.A. HAKIM PROGRAMMABLE VALVE WITH PRECHAMBER; SHUNT, CENTRAL NERVOUS SYSTEM & COMPS Back to Search Results
Catalog Number 82-3100
Device Problem Programming Issue (3014)
Patient Problem No Information (3190)
Event Type  Injury  
Manufacturer Narrative
(b)(4).Upon completion of the investigation, a follow up report will be filed.
 
Event Description
As reported by the ous affiliate, a hakim programmable valve could not be reprogrammed, so it was explanted.
 
Manufacturer Narrative
It was previously reported that the device would be returned for evaluation.Multiple attempts to obtain the sample were not successful.This report has been updated to reflect the corrected information.Complaint sample was not returned to codman and no lot number information has been provided; therefore, an evaluation of the device could not be performed and manufacturing records could not be reviewed.The cause(s) of the difficulty reported by the customer could not be determined.If the complaint sample becomes available, this complaint will be reopened, and the respective evaluation performed.Trends will be monitored for this or similar complaints.At present, we consider this complaint to be closed.
 
Manufacturer Narrative
Udi: (b)(4).The device was returned for evaluation.The position of the cam when valve was received was 40mmh2o.The valve was visually inspected: no defects noted.The valve was tested for programming and passed the test.The valve was flushed; no occlusion was noted.The valve was leak tested; no leaks noted.The valve was reflux tested and passed the test.The valve was then pressure tested; no issues found.A review of manufacturing records found the device conformed to specification when released to stock.Based on the results of the investigation, the reported issue could not be confirmed.The device functioned as intended.Trends will be monitored for this and similar issues.At present, we consider this complaint to be closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HAKIM PROGRAMMABLE VALVE WITH PRECHAMBER
Type of Device
SHUNT, CENTRAL NERVOUS SYSTEM & COMPS
Manufacturer (Section D)
CODMAN & SHURTLEFF, INC. / MEDOS S.A.
rue girardet 29
le locle CH 24 00
SZ  CH 2400
MDR Report Key7299247
MDR Text Key101057169
Report Number1226348-2018-10187
Device Sequence Number1
Product Code JXG
Combination Product (y/n)N
PMA/PMN Number
K974739
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/27/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date09/30/2022
Device Catalogue Number82-3100
Device Lot Number164249
Was Device Available for Evaluation? No
Date Manufacturer Received09/20/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-