• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATEX-OHMEDA, INC. AISYS CS2; ANESTHESIA GAS MACHINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATEX-OHMEDA, INC. AISYS CS2; ANESTHESIA GAS MACHINE Back to Search Results
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/19/2018
Event Type  malfunction  
Manufacturer Narrative
A gehc service representative performed a checkout of the equipment.The microswitch was replaced.Patient information could not be obtained after multiple attempts.Attempts were made as follows: 2/21/18 phone call, 2/26/18 phone call, 2/27/18 phone call.Unique identifier ¿ (b)(4).
 
Event Description
The hospital reported the bag/vent switch was not working as expected.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AISYS CS2
Type of Device
ANESTHESIA GAS MACHINE
Manufacturer (Section D)
DATEX-OHMEDA, INC.
3030 ohmeda dr,
madison, WI
Manufacturer Contact
john szalinski
3000 n grandview blvd.
waukesha, WI 
MDR Report Key7299739
MDR Text Key101084708
Report Number2112667-2018-00426
Device Sequence Number1
Product Code BSZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K132530
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 02/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/27/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Other Device ID NumberSEE H10
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received01/29/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/13/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-