| Model Number |
CSK-TC15 |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
01/30/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional suspect medical device component involved: model: csk-tc15 lot: m065 description: csk tc electrode, 15 cm.
|
| |
|
Event or Problem Description
|
|
A report was received that an electrode had a wire come out.
|
| |
|
Additional Manufacturer Narrative
|
|
Csk-tc15/lot l889: the complaint was confirmed.A visual inspection of both electrodes revealed that the shafts were bent due to external mechanical force.Both electrodes would also not register temperature during the functional test.X-ray inspection of the electrodes revealed that fractured wires in the shafts near the tips were causing an open thermocouple circuit.Csk-tc15/lot m678: the complaint was not confirmed.The electrode passed a visual inspection.However, the electrode would not register temperature during the functional test.X-ray inspection revealed that a fractured wire in the hub was causing an open thermocouple circuit.Csk-tc15/lot m039: the complaint was confirmed.A visual inspection of the electrode revealed that the shaft was bent.However, the electrode passed the functional test.
|
| |
|
Event or Problem Description
|
|
A report was received that an electrode had a wire come out.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional suspect medical device component involved: model: csk-tc15, lot: l889; description: csk tc electrode, 15 cm, total quantities: 2, model: csk-tc15, lot: m678; description: csk tc electrode, 15 cm, total quantities: 1, model: csk-tc15, lot: m039; description: csk tc electrode, 15 cm, total quantities: 1.
|
| |
|
Event or Problem Description
|
|
A report was received that an electrode had a wire come out.
|
| |
|