• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC. (LOUISVILLE) STEALTHSTATION S7 SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC. (LOUISVILLE) STEALTHSTATION S7 SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT Back to Search Results
Model Number S7
Device Problem Device Stops Intermittently (1599)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/01/2018
Event Type  malfunction  
Manufacturer Narrative
A medtronic representative went to the site to test the equipment.It was reported that the surgeon monitor for the navigation system was replaced to resolve the issue.The hardware, software, and instruments passed the system checkout.The system was found to be fully functional.The monitor for the navigation system was returned to the manufacturer for analysis.The device was found to be fully functional with no problem found.The reported event could not be duplicated by medtronic personnel.
 
Event Description
A medtronic representative reported that, while in a cranial resection, the surgeon monitor of the navigation system was "flickering" and that the monitor would intermittently lose functionality.A second monitor was used to complete the procedure.There was no reported delay to the procedure due to this issue.There was no impact on patient outcome.No additional information was provided.
 
Manufacturer Narrative
Correction: device manufacture date updated to proper value.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEALTHSTATION S7 SYSTEM
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
heather davis
attn:product quality experienc
826 coal creek circle
louisville, CO 80027-9710
MDR Report Key7299857
MDR Text Key101168479
Report Number1723170-2018-00901
Device Sequence Number1
Product Code HAW
UDI-Device Identifier00613994450944
UDI-Public00613994450944
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K050438
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup
Report Date 03/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/27/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model NumberS7
Device Catalogue Number9733856
Is the Reporter a Health Professional? No
Date Manufacturer Received03/05/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/17/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age61 YR
Patient Weight82
-
-