| Model Number |
2841401400 |
| Health Effect - Clinical Codes |
Hemorrhage/Blood Loss/Bleeding (1888); Urinary Tract Infection (2120); Injury (2348)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Without the benefit of examination and testing, coloplast is precluded from commenting on the condition of the device or the cause of the occurrence.Should additional facts prompt us to alter or supplement any information of conclusions contained in the original mdr or in any prior supplemental reports, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
According to the available information, the distributor submitted the following complaint on behalf of an end user.A patient complaint regarding item# 284140 - lot# 5976881.Patient stated that the 284140 catheters are not flexible enough and caused gauging and bleeding.No medical intervention reported.Answer provided by distributor: customer alleged that he injured himself with the catheter.He did not quantify the bleeding however he got a uti and is on antibiotics.The doctor placed him on a foley as well.The additional questions asked were: how much did the end user bleed? (was it a lot, just a little, we understand it is hard to judge but this information is important).Beside bleeding was there any other harm (i.E.Uti)? can you confirm no subsequent medical intervention was required since the complaint was filed.How long has the end user been using the product? eight years.How much did the end user bleed? only a little.So little the end user didn't know he was bleeding.His urologist told him.What was the end user's method of cathing? he followed directions exactly as instructed.End user has been using our cath #28414 for eight years.He had a problem with one of our caths which coincided with an infection.It may have had nothing to do with our cath.He became swollen from a bard cath.Their lubricating system wasn't good.Therefore our cath just didn't go through the last bend correctly and he used too much pressure.He bled very little.He didn't even know it, only because his urologist told him there is blood.
|
| |
|
Additional Manufacturer Narrative
|
|
The recheck of the relevant production documents did not reveal any deviation that could be related to the catheter is too stiff problem.We received 252 samples from the affected lot we made finger test on 20 catheters and we forwarded 5 samples for testing.All the performed tests were within specifications.This item was produced at (b)(4) site and the raw material was produced at (b)(4) site.Production was informed about the complaint.We continue to monitor the complaint trends and report it to the management regularly.
|
| |
|
Search Alerts/Recalls
|