| Model Number |
18885 |
| Medical Device Problem Code |
Inflation Problem (1310)
|
| Health Effect - Clinical Code |
Extubate (2402)
|
| Date of Event |
02/06/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during use of the devices, a cuff inflation/deflation issue occurred.The customer indicated that there w ere no patient symptoms or complications or harm associated with this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: the samples were received for evaluation and no failure was confirmed.No reported event was observed in the received samples since met with the product specifications.A visual inspection was performed and it was observed all the components are assembled properly.An inflation/deflation test was performed using a syringe of air was applied to the cuffs and it was noticed the cuffs held the air, the cuffs were left inflated 2 hours, after this time no deflation was observed, additionally the sample was submerged into a container filled with water and no leaks were observed.Information has been added to the database and trends will continue to be monitored.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during use of the devices, a cuff inflation/deflation issue occurred.There were no patient symptoms or complications though patient had to be reintubated.
|
| |
|
Search Alerts/Recalls
|