• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DUTCH OPTHALMIC RESEARCH CENTER BV EVA, COMBINED MACHINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DUTCH OPTHALMIC RESEARCH CENTER BV EVA, COMBINED MACHINE Back to Search Results
Model Number 8000.COM01
Medical Device Problem Codes Increase in Pressure (1491); Pressure Problem (3012)
Health Effect - Clinical Code Corneal Edema (1791)
Date of Event 02/02/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation has started.
 
Event or Problem Description
A clinic in spain has reported corneal edema after surgery.There is no further detailed information at this time.We have reached out for additional information and a follow up will be submitted should any new information be received.
 
Event or Problem Description
A report has been received from spain.The surgeon reported that during the eye surgery, the surgeon experienced high irrigation pressure into the eyes.Cornea edema appeared.A supplemental will be filed when new information is received.
 
Additional Manufacturer Narrative
Root cause investigation: 1.Dorc personnel has visited and observed the surgeon's operation, it was noticed that the customer uses very tight i / a irrigation flow, seems the pressure was high to irrigate the eye.2.Dorc have also reviewed the log files returned from the subject eva.It can be seen in the log files records that: only 21 surgeries were performed with the system.Vpgc and gravity modes were both set by the surgeons.The maximum irrigation pressure set by the surgeons was at 70 mmhg, which is not extremely high, and are generally used during surgery.However, the actual irrigation pressure into the air can be different from the pre-set value of eva system.The irrigation pressure could be too high if gravity is selected, while vgpc infusion accessories were connected, the later enables additional compressed air goes into the infusion bottle, while the infusion bottle was hang up at a relatively higher level, which will increase (accumulated up) the irrigation pressure higher than the preset pressure value than 70mmhg.Thus, to further investigate above assumption could be the root cause, we will check which packs were sent to the customer (only vgpc or also gravity packs of accessories).In a review the surgery experienced high irrigation pressure, most likely they were user operational errors, that the surgeons are not aware of how to use the eva gravity and vgpc modes.More investigation results will be reported in a next report.Dorc personnel has adjusted the pressure setting parameters.Instruct the surgeons on how to use the eva gravity and vgpc modes and pressure setting.
 
Event or Problem Description
New information has been received: the edema was reduced a few hours after the surgery.The surgeon did not prescribe any medication nor remedial interventions to the patient.
 
Additional Manufacturer Narrative
For the observed edema several root causes are feasible but in investigation of the unit log files no anomalies have been identified.Based upon questions to the customer, it was not possible to identify cases where vgpc kits were erroneously used whilst the unit was in gravity mode.The customer indicated that vgpc mode was used.Eva has been tested and is certified according to iec 80601-2-58.The design of the eva contains several safety measures related to the control of the pressure that is provided by the machine.Above a pressure setting of 60 mmhg the eva provides a visual and acoustic alert which can not be switched off.Log files of the surgeries concerned did not show any anomalies on the working of eva.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVA, COMBINED MACHINE
Common Device Name
EVA, COMBINED MACHINE
Manufacturer (Section D)
DUTCH OPTHALMIC RESEARCH CENTER BV
scheijdelveweg 2
zuidland, 3214V N
NL  3214VN
MDR Report Key7300272
Report Number1222074-2018-00036
Device Sequence Number11065797
Product Code HQC
UDI-Device Identifier08717872019642
UDI-Public08717872019642
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K142877
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 02/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Model Number8000.COM01
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/27/2018
Supplement Date Received by Manufacturer02/27/2018
04/18/2018
07/27/2018
Initial Report FDA Received Date02/27/2018
Supplement Report FDA Received Date02/27/2018
05/18/2018
07/27/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-