The patient was implanted with a left ventricular assist device (lvad) on (b)(6) 2017.It was reported that frequent low flow alarms occurred the morning of (b)(6) 2018.The patient was medically managed; however, the low flow alarms persisted.As part of troubleshooting, the system controller was exchanged; however, the change did not resolve the alarms.The lvad speed was manually decreased; however, the alarms persisted.On (b)(6) 2018, the patient continued to be asymptomatic.The patient's left ventricle had decreased in size by 3 cm.Because of this remodeling, the inflow cannula had likely changed position enough to cause the low flow alarms.No additional interventions were pursued, and the patient was discharged.
|