• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WARSAW ORTHOPEDICS VERTEX RECONSTRUCTION SYSTEM; APPLIANCE, FIXATION, SPINAL INTERLAMINAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WARSAW ORTHOPEDICS VERTEX RECONSTRUCTION SYSTEM; APPLIANCE, FIXATION, SPINAL INTERLAMINAL Back to Search Results
Catalog Number 7752529
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 02/03/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was not returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that patient underwent posterior cervical fusion at c3-c6 due to central canal stenosis of cervical spine with paralysis.Intra-op, set screw sheared.The mas cross link was placed over the male hex post and while placing the mas crosslink locking set nut, the male hex post sheared off.The residual screw remained in the mas vertex screw locking down the rod.Set screw was tightened using the torque and counter torque wrenches provided in the set.Because the rod was safely locked into the mas, no further steps were taken.The set screw remained in the patient; however the fragments were retrieved.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERTEX RECONSTRUCTION SYSTEM
Common Device Name
APPLIANCE, FIXATION, SPINAL INTERLAMINAL
Manufacturer (Section D)
WARSAW ORTHOPEDICS
2500 silveus crossing
warsaw IN 46582
MDR Report Key7300419
Report Number1030489-2018-00286
Device Sequence Number1029552
Product Code KWP
UDI-Device Identifier00613994454041
UDI-Public00613994454041
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
K082728
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number7752529
Device Lot Number0583931W
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/03/2018
Initial Report FDA Received Date02/27/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-