Model Number 8000.COM01
Medical Device Problem Codes Increase in Pressure (1491); Pressure Problem (3012)
Health Effect - Clinical Code Corneal Edema (1791)
Date of Event 02/02/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation has started.
 
Event or Problem Description
A report has been received from (b)(6).The surgeon reported that during eye surgery, the surgeon experienced high irrigation pressure into the eyes.Cornea edema appeared.No further information at this time.A request has been made for additional information.
 
Additional Manufacturer Narrative
Root cause investigation: 1.Dorc personnel has visited and observed the surgeon's operation, it was noticed that the customer uses very tight i / a irrigation flow, seems the pressure was high to irrigate the eye.2.Dorc have also reviewed the log files returned from the subject eva.It can be seen in the log files records that: _ only 21 surgeries were performed with the system._ vpgc and gravity modes were both set by the surgeons._ the maximum irrigation pressure set by the surgeons was at 70 mmhg, which is not extremely high, and are generally used during surgery.However, the actual irrigation pressure into the air can be different from the pre-set value of eva system.The irrigation pressure could be too high if gravity is selected, while vgpc infusion accessories were connected, the later enables additional compressed air goes into the infusion bottle, while the infusion bottle was hang up at a relatively higher level, which will increase (accumulated up) the irrigation pressure higher than the preset pressure value than 70mmhg.Thus, to further investigate above assumption could be the root cause, we will check which packs were sent to the customer (only vgpc or also gravity packs of accessories).In a review the surgery experienced high irrigation pressure, most likely they were user operational errors, that the surgeons are not aware of how to use the eva gravity and vgpc modes.More investigation results will be reported in a next report.Dorc personnel has adjusted the pressure setting parameters.Instruct the surgeons on how to use the eva gravity and vgpc modes and pressure setting.
 
Event or Problem Description
New information has been received: the edema was reduced a few hours after the surgery.The surgeon did not prescribe any medication nor remedial interventions to the patient.
 
Additional Manufacturer Narrative
Root cause investigation: 1.Dorc personnel has visited and observed the surgeon's operation, it was noticed that the customer uses very tight i / a irrigation flow, seems the pressure was high to irrigate the eye.2.Dorc have also reviewed the log files returned from the subject eva.It can be seen in the log files records that: _ only 21 surgeries were performed with the system._ vpgc and gravity modes were both set by the surgeons._ the maximum irrigation pressure set by the surgeons was at 70 mmhg, which is not extremely high, and are generally used during surgery.However, the actual irrigation pressure into the air can be different from the pre-set value of eva system.The irrigation pressure could be too high if gravity is selected, while vgpc infusion accessories were connected, the later enables additional compressed air goes into the infusion bottle, while the infusion bottle was hang up at a relatively higher level, which will increase (accumulated up) the irrigation pressure higher than the preset pressure value than 70mmhg.Thus, to further investigate above assumption could be the root cause, we will check which packs were sent to the customer (only vgpc or also gravity packs of accessories).In a review the surgery experienced high irrigation pressure, most likely they were user operational errors, that the surgeons are not aware of how to use the eva gravity and vgpc modes.More investigation results will be reported in a next report.Dorc personnel has adjusted the pressure setting parameters.Instruct the surgeons on how to use the eva gravity and vgpc modes and pressure setting.
 
Event or Problem Description
New information has been received: the edema was reduced a few hours after the surgery.The surgeon did not prescribe any medication nor remedial interventions to the patient.
 
Additional Manufacturer Narrative
For the observed edema several root causes are feasible but in investigation of the unit log files no anomalies have been identified.Based upon questions to the customer, it was not possible to identify cases where vgpc kits were erroneously used whilst the unit was in gravity mode.The customer indicated that vgpc mode was used.Eva has been tested and is certified according to iec 80601-2-58.The design of the eva contains several safety measures related to the control of the pressure that is provided by the machine.Above a pressure setting of 60 mmhg the eva provides a visual and acoustic alert which can not be switched off.Log files of the surgeries concerned did not show any anomalies on the working of eva.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVA, COMBINED MACHINE INCLUDING LASER ( DORC CONNECTOR)
Common Device Name
PHACOEMULSIFICATION/VITRECTOMY SYSTEM
Manufacturer (Section D)
DUTCH OPTHALMIC RESEARCH CENTER BV
scheijdelveweg 2
zuidland, 3214V N
NL  3214VN
Manufacturer (Section G)
DUTCH OPTHALMIC RESEARCH CENTER BV
scheijdelveweg 2
zuidland, 3214V N
NL   3214VN
Manufacturer Contact
laura smith
10 continental dr
bldg 1
exeter, NH 03833
6037786929
MDR Report Key7300517
Report Number1222074-2018-00037
Device Sequence Number10986541
Product Code HQC
UDI-Device Identifier08717872029672
UDI-Public08717872029672
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K142877
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number8000.COM01
Device Catalogue Number8000.COM01
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/06/2018
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date02/27/2018
Supplement Report FDA Received Date02/04/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/27/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-