Model Number 7211010S
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the pump was running higher than it should, the flow rate kept jumping up.No patient injury or significant time delay reported.
 
Additional Manufacturer Narrative
The device was received for evaluation.There was no relationship found between the returned device and the reported incident.A visual inspection was performed on the product and no issue was observed.Complaint of erratic flow rate was not confirmed.Product passed 2 hours on the wet test station but failed functional testing with transducer p1 reading of 480mmhg in the 500mmhg pressure test.Specs are 500 +/- 10mmhg.Also transducer p2 failed raw output test with 18.5mmhg.Specs are 0.0 +/- 15mmhg.Cause of failures are defective transducers p1 and p2.Product passed functional testing with two known good transducers installed.The complaint of erratic flow rate was not confirmed and the root cause could not be determined since the reported malfunction could not be duplicated during the product evaluation process.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SVCE REPL CONTROL UNIT DYONICS 25
Common Device Name
ARTHROSCOPE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
MDR Report Key7300583
Report Number1643264-2018-00175
Device Sequence Number1276462
Product Code HRX
UDI-Device Identifier03596010555960
UDI-Public(01)03596010555960
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K051326
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number7211010S
Device Catalogue Number7211010S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/12/2018
Initial Date Received by Manufacturer 02/23/2018
Supplement Date Received by Manufacturer04/18/2018
Initial Report FDA Received Date02/27/2018
Supplement Report FDA Received Date04/19/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-