| Model Number |
VT1737 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
02/07/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during ablation, the device broke when placing it on the ribs.New device was used to complete the case.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary one device was received for evaluation.This device had been used in the treatment or diagnosis of a patient.A review of the lot number reported indicates that the product was within the assigned expiration date at the time of the reported incident.The returned product did not meet specification as received.Visual inspection found the antenna was broken at the plastic feed point.The customer reported that the device broke when placing it on the ribs.The reported condition was confirmed.The investigation found that the plastic feedpoint in the radiating section was fractured.The inner conductor wire is still intact.The investigation identified the root cause of the reported event to be user error.The instruction for use states: do not apply excessive lateral or rotational force during insertion or extraction of the antenna.Doing so may lead to breakage at the antenna feedpoint and injury to the patient or user.Use a scalpel or electrosurgical pencil to expand the insertion point if you meet resistance.Although the antenna is slightly flexible, using force may lead to breakage at the antenna feedpoint and possible injury to the patient.Do not attempt to bend or reshape antennas.Product damage and/or malfunction may result.View the inserted portion of the antenna using image guidance, such as ct or ultrasound, to ensure the antenna is not bent or broken.No update to the risk management file is required at this time.Manufacturing non-conformance were reviewed.No entries pertinent to the customer's report were noted.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during ablation, the device broke when placing it on the ribs; however, the doctor was able to retract the antennas without damage.No part of the antenna came loose; it just fell over and the antenna was salvaged successfully.New device was used to complete the case.There was no patient injury.
|
| |
|
Search Alerts/Recalls
|