• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. ULTRAVERSE 035 PTA BALLOON DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. ULTRAVERSE 035 PTA BALLOON DILATATION CATHETER Back to Search Results
Model Number U35130530
Medical Device Problem Codes Deflation Problem (1149); Improper or Incorrect Procedure or Method (2017)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 02/01/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No medical records and no medical images were provided to the manufacturer.The lot number for the device was provided.The device history records are currently under review.The return of the device is pending.The investigation is currently under way.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that after the first inflation, the pta balloon catheter allegedly would not deflate.It was further reported that the health care provider used an inflation device to over inflate the balloon and cause it to rupture.Reportedly, the balloon catheter was removed without incident.No further treatment was provided.There was no reported patient injury.
 
Additional Manufacturer Narrative
Manufacturing review: the lot number was provided and the lot device history records have been reviewed.The lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.This is the only event reported to date for this lot number and failure mode.Investigation summary: the device was returned.A visual inspection found a longitudinal rupture extending for the majority of the balloon length.Per the reported event details, the user over-inflated the device to rupture the balloon, as the device allegedly could not be deflated.Therefore, the investigation is confirmed for a use-of-device issue.The investigation is inconclusive for the reported deflation issue, as the device was unable to be functionally tested due to the longitudinal rupture.The definitive root cause for the reported deflation issue could not be determined based upon the available information.Per the reported event details, the user over-inflated the device to rupture the balloon, as the balloon could allegedly not be deflated after dilatation.The current instructions for use states to not exceed the rbp recommended for the device, and to remove the balloon catheter and introducer sheath/guide sheath as a single unit, if resistance is felt during post procedure withdrawal of the device.Labeling review: the current ifu (instructions for use) states: warnings: do not exceed the rbp recommended for this device.Balloon rupture may occur if the rbp rating is exceeded.To prevent over pressurization, use of a pressure monitoring device is recommended.Precautions: if resistance is still felt during post procedure withdrawal of the catheter, it is recommended to remove the balloon catheter and introducer sheath/guide catheter is a single unit.Use of the ultraverse 035 pta dilatation catheter: position the balloon relative to the lesion to be dilated, ensure the guidewire is in place, and inflate the balloon to the appropriate pressure.
 
Event or Problem Description
It was reported that after the first inflation, the pta balloon catheter allegedly would not deflate.It was further reported that the health care provider used an inflation device to over inflate the balloon and cause it to rupture.Reportedly, the balloon catheter was removed without incident.No further treatment was provided.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRAVERSE 035 PTA BALLOON DILATATION CATHETER
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key7301092
Report Number2020394-2018-00165
Device Sequence Number13755962
Product Code LIT
UDI-Device Identifier00801741092756
UDI-Public(01)00801741092756
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K142261
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2020
Device Model NumberU35130530
Device Catalogue NumberU35130530
Device Lot NumberCMBY0109
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/06/2018
Initial Date Received by Manufacturer 02/01/2018
Supplement Date Received by Manufacturer05/01/2018
Initial Report FDA Received Date02/27/2018
Supplement Report FDA Received Date05/11/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age74 YR
Patient Weight81
-
-