MDT SOFAMOR DANEK PUERTO RICO MFG CD HORIZON SPINAL SYSTEM; APPLIANCE, FIXATION, SPINAL INTERLAMINAL Back to Search Results
Catalog Number 75446540
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 12/27/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Image review: post-op x-ray, l4-s1 fusion repeated as l3-l5.2017 xrays show fracture of s1 screws with a portion of the sq screw retained in the sacrum.There is a paucity of bone in the lateral spine and an interbody graft was used.Product was not returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Levels implanted: l3-l5 it was reported that the patient presented with central stenosis and instability and underwent stabilization with screws.Post-op, the screw broke.Revision surgery was performed to remove the screw.Part of the screw remained in patient.
 
Additional Manufacturer Narrative
Visual and microscopic examination of mas bone screw confirms breakage approximately 3 threads from the base of the bone screw head, microscopic examination reveals major smearing of the fracture surface.There is little to no area of the fracture face that is not damaged.The witness marks in the saddle of the mas are not atypical and do not point to an installation issue.At this time due to the damage to the fracture face, a root cause cannot be determined.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CD HORIZON SPINAL SYSTEM
Common Device Name
APPLIANCE, FIXATION, SPINAL INTERLAMINAL
Manufacturer (Section D)
MDT SOFAMOR DANEK PUERTO RICO MFG
barrio marianna rd 909, km0.4
humacao PR 00792
MDR Report Key7302682
Report Number1030489-2018-00295
Device Sequence Number11341365
Product Code KWP
UDI-Device Identifier00885074134116
UDI-Public00885074134116
Combination Product (Y/N)N
Initial Reporter CountryUP
PMA/510(K) Number
K042025
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number75446540
Device Lot NumberH10J1956
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/27/2018
Initial Date Received by Manufacturer 01/29/2018
Supplement Date Received by Manufacturer04/27/2018
Initial Report FDA Received Date02/28/2018
Supplement Report FDA Received Date05/02/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-