Consumer reported complaint for high blood glucose test results.The customer is concerned with tests results from results obtained of 246 and 272 mg/dl.The customer's expected fasting blood glucose test result range is 125 - 160 mg/dl.The customer feels well and did not report any symptoms.Medical attention is not reported as a result of the actual blood glucose results.During the call on (b)(6) 018, a back to back blood test was not performed by the customer.The product is not stored according to specification and is stored in the kitchen.The test strip lot manufacturer's expiration date is 10/31/2018 and open vial date at time of call is three weeks.The meter memory was reviewed for previous test result history: result 1 :158mg/dl date (b)(6) 2018 time:6:50pm fasting, result 2 :246mg/dl date: (b)(6) 2018 time:2:50 pm fasting, result 3 :272mg/dl date: (b)(6) 2018 time:10:24am fasting, customer states that his meter started testing high within the last few weeks.Customer compared the results based on how he felt at the time of testing.The results of concerned are 246 and 272mg/dl fasting.Expected range is 125-160mg/dl fasting.Customer states that he does not want his meter replaced, he only wants to order control solutions to test if the meter is defective.
|