| Model Number |
M00565010 |
| Medical Device Problem Code |
Premature Activation (1484)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
02/05/2016
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The device has been received for analysis; however the evaluation has not been completed.Therefore, the cause of the reported malfunction has not been determined.Upon completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation on (b)(6) 2018 that a wallflex enteral duodenal stent was to be used during an endoscopic retrograde cholangiopancreatography (ercp) with duodenal stent placement procedure performed on (b)(6) 2016.According to the complainant, during the procedure, the stent was prematurely deployed outside the scope.The procedure was completed with a different size of wallflex enteral duodenal stent.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be fine.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation on (b)(6) 2018 that a wallflex enteral duodenal stent was to be used during an endoscopic retrograde cholangiopancreatography (ercp) with duodenal stent placement procedure performed on (b)(6) 2016.According to the complainant, during the procedure, the stent was prematurely deployed outside the scope.The procedure was completed with a different size of wallflex enteral duodenal stent.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be fine.
|
| |
|
Additional Manufacturer Narrative
|
|
A fully deployed wallflex duodenal stent and delivery system were received for analysis.Visual examination of the returned device found the stainless steel tube was severely kinked.The stent was measured to be within specifications.There were no issues noted with the stent.No other issues with the device were noted during the product analysis.The damage noted with the device was consistent with the application of excessive force being applied to the device.The investigation concluded that the observed failure and the reported event were likely due to anatomical or procedural factors encountered during the procedure which limited the performance of the device.Therefore, the most probable root cause is operational context.A review of the device history record (dhr) confirmed that the device met all material, assembly, and product specifications at the time of release to distribution.A search of the complaint database revealed that no other similar complaints exist for the specified lot.
|
| |
|