| Catalog Number |
0066814 |
| Medical Device Problem Code |
False Negative Result (1225)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
02/05/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The immucor laboratory tested retention product on (b)(6) 2018, which performed as expected.The immucor internal identification number for this report is (b)(4).
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2018, a (b)(6) customer site reported an unexpected negative antibody screen when using capture-r ready-screen (3) with a galileo neo instrument.
|
| |
|
Search Alerts/Recalls
|