| Catalog Number |
8670087 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
02/02/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device was not returned to manufacturer for evaluation therefore we are unable to determine the definitive cause of event.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that patient presented with degenerative disc disease underwent lumbar decompression and fusion(minimally invasive approach) at l3-s1.Intra-op, tip of drill tap broke off.Tap immediately removed from surgical field and was replaced with another.No patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Product analysis result: visual microscopic hardness an approx.1.5mm section has broken off the tip of the tap.A microscopic review of the fracture face reveals a brittle fracture with rays emanating outward.This type of fracture is consistent with the material being overloaded by coming to a sudden stop during rotation.The hardness of the material was checked and meets print spec.The observations are consistent with torsional overload.If information is provided in the future, a supplemental report will be issued.
|
| |
|