| Model Number |
72202797 |
| Health Effect - Clinical Code |
Device Embedded In Tissue or Plaque (3165)
|
| Date of Event |
02/22/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that it was used a 1.2mm guide wire to guide the biorci screw and once screw was implanted he tried to remove guidewire that appeared to be stuck.Guidewire broke leaving part of wire in the bone.It was took an x-ray post operation and it was saw that approx.5mm was in the bone.It was left in the patient as is.
|
| |
|
Additional Manufacturer Narrative
|
|
One 1.2mm nitinol guidewire from a rap-pak b was returned for evaluation.Visual assessment of the guidewire confirmed the reported breakage.Approximately 3/8¿ of the distal end of the guidewire has been broken off.The broken piece was not returned.The break area is bent and shows signs of elastic deformation.The devices condition indicates it was subjected to excessive force during use.As with any surgical instrument, careful attention should be made to assure that excessive force is not placed on the instrument.Excessive forces applied to the instrument can result in failure.No root cause related to the manufacture of the device can be established.
|
| |
|