510(k) number: k142688.Cook ireland ltd (manufacturer) is submitting this report on behalf of cook medical incorporated (cmi)(importer).Exemption number: e2016031.Importer site contact and address: (b)(6).Importer site establishment registration number: (b)(4).Problem statement: ¿ echotip ultra 25 ga needle bent after extending from sheath.Product available for return and inspection.Lot number c1404483.¿ additional information was requested the 20th feb 2018:"do you have answers for the additional questions as the device has not returned we need as much information from the customer as possible." lab evaluation: it was originally indicated that the device involved in this complaint was being returned to cook ireland for evaluation; the device has not yet been received, however, if it is returned, the device will be evaluated and the investigation will be updated.As the device was not returned for evaluation; the cause of this complaint could not be conclusively determined.With the information provided a document based investigation was carried out.The customer complaint is considered to be confirmed based on customer testimony.Root cause: a definitive root cause for the customer complaint could not be determined, however the device may have been placed under extreme pressure which caused the needle to bend.Prior to distribution, all echo devices are subjected to functional checks and visual inspection to ensure device integrity.These inspections and functional checks are outlined in internal procedures in place at cirl.(b)(4).All products and packaging are 100% inspected for visual irregularities such as holes, dents, kinks or tears there is a check on each device to measure the needle extension and inspect it to ensure it is free from any damage / kinks.There is also a check in place to ensure that the stylet wire operates smoothly.Verify the needle fully retracts into the sheath.As instructed in fqc ¿check for visual defects, debris, loose or embedded foreign materials, rough or sharp edges, bends or kinks.¿ a review of the manufacturing records did not reveal any discrepancies that would have contributed to the complaint.There is no evidence to suggest that this issue affects the entire lot #; upon review of complaints this failure mode has not occurred previously with this lot #.The notes section of the instructions for use that accompanies this device instructs the user to inspect the device prior to use for any damage: "visually inspect with particular attention to kinks, bends and breaks.If an abnormality is detected that would prohibit proper working condition, do not use".The tip of the needle and stylet are sharp and could cause injury to the patient or user if not used with caution.On review of the information provided, there is no viable evidence to suggest that the user did not follow the instructions for use.From the information provided: the patient did not experience any adverse effects due to this occurrence.Complaints of this nature will continue to be monitored for potential emerging trends.510(k) number: k142688.Cook ireland ltd (manufacturer) is submitting this report on behalf of cook medical incorporated (cmi)(importer).Exemption number: e2016031.Importer site contact and address: (b)(6) importer site establishment registration number: (b)(4).Problem statement: ¿ echotip ultra 25 ga needle bent after extending from sheath.Product available for return and inspection.Lot number c1404483.¿ additional information was requested the 20th feb 2018:"do you have answers for the additional questions as the device has not returned we need as much information from the customer as possible." lab evaluation: it was originally indicated that the device involved in this complaint was being returned to cook ireland for evaluation; the device has not yet been received, however, if it is returned, the device will be evaluated and the investigation will be updated.As the device was not returned for evaluation; the cause of this complaint could not be conclusively determined.With the information provided a document based investigation was carried out.The customer complaint is considered to be confirmed based on customer testimony.Root cause: a definitive root cause for the customer complaint could not be determined, however the device may have been placed under extreme pressure which caused the needle to bend.Prior to distribution, all echo devices are subjected to functional checks and visual inspection to ensure device integrity.These inspections and functional checks are outlined in internal procedures in place at cirl.(b)(4).All products and packaging are 100% inspected for visual irregularities such as holes, dents, kinks or tears there is a check on each device to measure the needle extension and inspect it to ensure it is free from any damage / kinks.There is also a check in place to ensure that the stylet wire operates smoothly.Verify the needle fully retracts into the sheath.As instructed in fqc ¿check for visual defects, debris, loose or embedded foreign materials, rough or sharp edges, bends or kinks.¿ a review of the manufacturing records did not reveal any discrepancies that would have contributed to the complaint.There is no evidence to suggest that this issue affects the entire lot #; upon review of complaints this failure mode has not occurred previously with this lot #.The notes section of the instructions for use that accompanies this device instructs the user to inspect the device prior to use for any damage: "visually inspect with particular attention to kinks, bends and breaks.If an abnormality is detected that would prohibit proper working condition, do not use".The tip of the needle and stylet are sharp and could cause injury to the patient or user if not used with caution.On review of the information provided, there is no viable evidence to suggest that the user did not follow the instructions for use.From the information provided: the patient did not experience any adverse effects due to this occurrence.Complaints of this nature will continue to be monitored for potential emerging trends.
|