| Model Number |
G25670 |
| Medical Device Problem Code |
Device Markings/Labelling Problem (2911)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
02/01/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Cook ireland ltd (manufacturer) is submitting this report on behalf of cook medical incorporated (cmi)(importer).Exemption number: e2016031.Information pertaining to section g.1 as follows: importer site contact and address: (b)(4) cook medical incorporated (cmi) 1025 acuff road p.O box 4195 bloomington indiana 47402-4195.Importer site establishment registration number: 3005580113.Investigation is still pending.A follow up mdr will be submitted to include the investigation conclusions.
|
| |
|
Event or Problem Description
|
|
As reported to customer relations, "the radiopaque markers came off the solus stent during insertion.The physician left the stent in place and the markers fell inside the stomach.The physician sprayed the introducer with silicone before hand.".
|
| |
|
Event or Problem Description
|
|
As reported to customer relations, "the radiopaque markers came off the solus stent during insertion.The physician left the stent in place and the markers fell inside the stomach.The physician sprayed the introducer with silicone before hand.".
|
| |
|
Additional Manufacturer Narrative
|
|
Cook ireland ltd (manufacturer) is submitting this report on behalf of cook medical incorporated (cmi)(importer).Exemption number: e2016031 information pertaining to mfr site as follows: importer site contact and address: (b)(6).(b)(4).Additional information 21 feb 2018; 1.Did the physician place the stent in the biliary duct to drain into the duodenum or was the stent placed somewhere else to drain into the stomach? ¿i was doing a transgastric pseudocyst drainage.¿ 2.Could you confirm what model and make of echo endoscope was used? ¿the scope was a pentax eg-3870utk 3.8.¿ lab evaluation: the complaint device was not returned to cirl for evaluation.This complaint is confirmed based on customer testimony.Root cause: a definitive root cause for the customer complaint could not be determined as the exact operational conditions of use could not be replicated in the laboratory setting.However it is possible that the bands came off as a result of the user choosing to drain through the stomach rather than through the biliary duct using the duodenal approach (trans-sphincter).This is considered to be 'off- label use'.Insertion of the stent through the stomach wall would be a more restricted route compared to placement in the biliary duct.This may have forced the bands down the stent and thus fall into the stomach.Ifu review: it may be noted that according to the instructions for use, ifu0100-1, the user is instructed to: ¿visually inspect with particular attention to kinks, bends and breaks.If any abnormality is detected that would prohibit proper working condition, do not use¿.As per the intended use section of the ifu; this device is used to drain obstructed biliary ducts." however, in this case the device was used for trans- gastric drainage of a pseudocyst.Fqc/ prd review: prior to distribution, all zss devices are subjected to 100% inspection to ensure device integrity.These inspections and functional checks are outlined in internal procedures in place at cirl.A review of the manufacturing records for the zss device of lot: c1361038 did not reveal any discrepancy related to the complaint issue.There is no evidence to suggest that this issue affects the entire lot#: c1361038; upon review of complaints this failure mode has not occurred previously with this lot#: c1361038.Summary: this complaint is confirmed based on customer testimony.Complaints of this nature will continue to be monitored for potential emerging trends.
|
| |
|
Additional Manufacturer Narrative
|
|
Cook ireland ltd (manufacturer) is submitting this report on behalf of cook medical incorporated (cmi)(importer).Exemption number: e2016031.Information pertaining to section g.1 as follows: importer site contact and address: (b)(6).Importer site establishment registration number: 3005580113.Additional information 21 feb 2018; 1.Did the physician place the stent in the biliary duct to drain into the duodenum or was the stent placed somewhere else to drain into the stomach? ¿i was doing a transgastric pseudocyst drainage.¿ 2.Could you confirm what model and make of echo endoscope was used? ¿the scope was a pentax eg-3870utk 3.8.¿ lab evaluation: the complaint device was not returned to cirl for evaluation.This complaint is confirmed based on customer testimony.Root cause: a definitive root cause for the customer complaint could not be determined as the exact operational conditions of use could not be replicated in the laboratory setting.However it is possible that the bands came off as a result of the user choosing to drain through the stomach rather than through the biliary duct using the duodenal approach (trans-sphincter).This is considered to be 'off- label use'.Insertion of the stent through the stomach wall would be a more restricted route compared to placement in the biliary duct.This may have forced the bands down the stent and thus fall into the stomach.Ifu review: it may be noted that according to the instructions for use, ifu0100-1, the user is instructed to: ¿visually inspect with particular attention to kinks, bends and breaks.If any abnormality is detected that would prohibit proper working condition, do not use¿.As per the intended use section of the ifu; this device is used to drain obstructed biliary ducts." however, in this case the device was used for trans- gastric drainage of a pseudocyst.Fqc/ prd review: prior to distribution, all zss devices are subjected to 100% inspection to ensure device integrity.These inspections and functional checks are outlined in internal procedures in place at cirl.A review of the manufacturing records for the zss device of lot c1361038 did not reveal any discrepancy related to the complaint issue.There is no evidence to suggest that this issue affects the entire lot # c1361038; upon review of complaints this failure mode has not occurred previously with this lot # c1361038.Summary: this complaint is confirmed based on customer testimony.Complaints of this nature will continue to be monitored for potential emerging trends.H3 other text : cook ireland ltd (manufacturer) is submitting this report on behalf of cook medical incorporated (cmi)(importer).Exemption number: e2016031 information pertaining to section g.1 as follows: importer site contact and address: (b)(6).Importer site establishment registration number: 3005580113 additional information 21 feb 2018; 1.Did the physician place the stent in the biliary duct to drain into the duodenum or was the stent placed somewhere else to drain into the stomach? ¿i was doing a transgastric pseudocyst drainage.¿ 2.Could you confirm what model and make of echo endoscope was used? ¿the scope was a pentax eg-3870utk 3.8.¿ lab evaluation: the complaint device was not returned to cirl for evaluation.This complaint is confirmed based on customer testimony.Root cause: a definitive root cause for the customer complaint could not be determined as the exact operational conditions of use could not be replicated in the laboratory setting.However it is possible that the bands came off as a result of the user choosing to drain through the stomach rather than through the biliary duct using the duodenal approach (trans-sphincter).This is considered to be 'off- label use'.Insertion of the stent through the stomach wall would be a more restricted route compared to placement in the biliary duct.This may have forced the bands down the stent and thus fall into the stomach.Ifu review: it may be noted that according to the instructions for use, ifu0100-1, the user is instructed to: ¿visually inspect with particular attention to kinks, bends and breaks.If any abnormality is detected that would prohibit proper working condition, do not use¿.As per the intended use section of the ifu; this device is used to drain obstructed biliary ducts." however, in this case the device was used for trans- gastric drainage of a pseudocyst.Fqc/ prd review: prior to distribution, all zss devices are subjected to 100% inspection to ensure device integrity.These inspections and functional checks are outlined in internal procedures in place at cirl.A review of the manufacturing records for the zss device of lot c1361038 did not reveal any discrepancy related to the complaint issue.There is no evidence to suggest that this issue affects the entire lot # c1361038; upon review of complaints this failure mode has not occurred previously with this lot # c1361038.Summary: this complaint is confirmed based on customer testimony.Complaints of this nature will continue to be monitored for potential emerging trends.
|
| |
|
Event or Problem Description
|
|
This mdr is being submitted as a correction, based on additional information and clinical input received there was no harm to the patient.This report has been re-categorized as a malfunction report only.Confirmation was received that the markers fell into the stomach and were excreted from the body naturally.Clinical input was received (b)(6) 2018: would you be able to give your clinical input on the potential harm to the patient by excreting the radiopaque markers? there is no harm if the markers were expelled naturally.As reported to customer relations, "the radiopaque markers came off the solus stent during insertion.The physician left the stent in place and the markers fell inside the stomach.The physician sprayed the introducer with silicone before hand.".
|
| |
|