• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: FLOWONIX MEDICAL, INC. PROMETRA INTRATHECAL CATHETER; IMPLANTABLE INTRATHECAL CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FLOWONIX MEDICAL, INC. PROMETRA INTRATHECAL CATHETER; IMPLANTABLE INTRATHECAL CATHETER Back to Search Results
Model Number 11823
Device Problem Device Issue (2379)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/08/2018
Event Type  Injury  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that the volume of undelivered drug remaining in the drug reservoir was greater than the expected volume based upon the programmed infusion rate.The physician was unable to aspirate fluid from the catheter access port (cap).Surgical intervention was taken to revise the catheter.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROMETRA INTRATHECAL CATHETER
Type of Device
IMPLANTABLE INTRATHECAL CATHETER
Manufacturer (Section D)
FLOWONIX MEDICAL, INC.
500 international drive
suite 200
mount olive NJ 07828
Manufacturer Contact
karen matis
500 international drive
suite 200
mount olive, NJ 07828
9734269229
MDR Report Key7303761
MDR Text Key101237556
Report Number3010079947-2018-00021
Device Sequence Number1
Product Code LKK
UDI-Device Identifier00810335020099
UDI-Public0100810335020099171812281022894
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P080012
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Physician
Type of Report Initial
Report Date 12/04/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Physician
Device Expiration Date12/28/2018
Device Model Number11823
Device Catalogue Number11823
Device Lot Number22894
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/31/2018
Initial Date FDA Received02/28/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/28/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-