Model Number QUANTUM 600 |
Device Problem
Fire (1245)
|
Patient Problem
No Code Available (3191)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The "date of event" has not been provided.The device has not been made available for evaluation at this time.Should further information or the device(s) become available, a follow-up report will be generated.
|
|
Event Description
|
Alleges chair caught on fire due to batteries arcing.Alleges user had to spray unit down with fire extinguisher.
|
|
Manufacturer Narrative
|
The device was returned and evaluated.The evaluation concluded there was an exposed battery terminal.
|
|
Event Description
|
Alleges chair caught on fire due to batteries arcing.Alleges user had to spray unit down with fire extinguisher.
|
|