| Catalog Number |
860284 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
The following instruments didn¿t function correctly during knee revision surgery.The 860284.Lot unknown -pfc mod plus torque driver didn¿t appear to click at maximum torque causing the wobble bits to fracture.The 961673.Lot so2008134.-adapter torque wrench.The black plastic adapter that positions over the femoral adapter detached from the instrument during tightening.The 217863136.Lot tbjgf.- modified hudson adapter got stuck on one of the mbt reamers.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint #: (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional narrative: product complaint # (b)(4).Investigation summary: examination of the returned instrument confirmed the reported event.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot: null.Device history batch: null.Device history review: null.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|