(b)(4).The device was not returned for analysis.It should be noted that in the supera instructions for use (ifu), the recommended pre-dilatation diameter for a 6.0mm stent is to use a balloon diameter greater than 6.0mm.It is possible that failing to pre-dilate the lesion site as instructed in the ifu contributed to the reported difficulties.With the vessel diameter being smaller than 6.0mm to accommodate the stent diameter, the stent would likely have elongated likely preventing the stent from fully releasing from the delivery system; however, this could not be confirmed.A review of the lot history record revealed no manufacturing nonconformities that would have contributed to this complaint.Additionally, a review of the complaint history of the reported lot revealed no other incidents.The investigation was unable to determine a conclusive cause for the deployment issue.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
|