| Model Number |
TRUEMETRIX |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
02/02/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Internal report # (b)(4).Product not yet returned for evaluation.Most likely underlying root cause: mlc-58-user had an inaccurate reference: self: the person is using themselves as the reference or how they feel at the time they run the blood test.Test strip udi# (b)(4).Note: manufacturer contacted customer (several attempts) in a follow-up call in order to ensure that the replacement products resolved the initial concern - unable to establish contact with the customer at this time.Product notification letter sent to customer to contact customer care.
|
| |
|
Event or Problem Description
|
|
Consumer reported complaint for high blood glucose test results.The customer is concerned with tests results from results of 180mg/dl.The customer's expected fasting blood glucose test result range is 100 - 120 mg/dl.The customer feels well and did not report any symptoms.Medical attention is not reported as a result of the actual blood glucose results.During the call on (b)(6) 2018, a back to back blood test was not performed by the customer; customer stated she was at work and did not have time.The product is stored according to specification in the bedroom.The test strip lot manufacturer's expiration date is 02/28/2019 and open vial date is january 2018.The meter memory was reviewed for previous test result history: (b)(6).
|
| |
|
Search Alerts/Recalls
|