MAUDE Adverse Event Report: AV-TEMECULA-CT OMNILINK ELITE PERIPHERAL STENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT OMNILINK ELITE PERIPHERAL STENT SYSTEM Back to Search Results
Catalog Number 1012631-19
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 02/21/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Internal file number - (b)(4): during processing of this complaint, attempts were made to obtain complete event, patient and device information.The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history of the reported lot revealed no other incidents.The investigation determined the reported difficulties appear to be related to circumstances of the procedure as it is likely that as the sds was advanced interaction with the heavily calcified anatomy and/or other devices resulted in the reported stent dislodgement.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Event or Problem Description
It was reported the procedure was to treat the heavily calcified left and right iliac arteries with a kissing stent technique at the bifurcation.The 8.0x19mm otw omnilink elite stent delivery system (sds) was advanced to the right iliac while another sds was advanced to the left iliac simultaneously.Once the omnilink elite sds was placed in the target lesion, it was noted the stent was not on the balloon, but had dislodged proximally onto the shaft.Negative was pulled on the balloon and the sds was retracted until the stent was over the balloon.The sds was then advanced to the target lesion and the stent was deployed successfully.The physician stated he did not know whether the stent caught on the 6fr sheath or interacted with the anatomy during advancement.There was no clinically significant delay in the procedure and no adverse patient effects.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OMNILINK ELITE PERIPHERAL STENT SYSTEM
Common Device Name
PERIPHERAL STENT SYSTEM
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key7304687
Report Number2024168-2018-01460
Device Sequence Number13791808
Product Code NIO
UDI-Device Identifier08717648178498
UDI-Public08717648178498
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
P110043
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date (Section B) 02/28/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2019
Device Catalogue Number1012631-19
Device Lot Number6062241
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/21/2018
Initial Report FDA Received Date02/28/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age70 YR
Patient Weight82
-
-